Office: Almería, Spainsourcing@alisios.co+34 655 099 909
Reference · Emerson Rosemount

How to read a Rosemount 3051 model code

A Rosemount 3051C model code is a string of fixed positions followed by option codes. Each position fixes one property of the transmitter. This page lists the positions and their common codes as published in the Emerson product data sheet for the Rosemount 3051, and decodes one example.

Updated: November 2026

Request for quotation
Send the part number or a photo of the nameplate.
Reply within 4 business hours.

Quotations are free and carry no obligation. * required.
We use your details only to answer this request. Privacy.
sourcing@alisios.co ·

What does each position in a Rosemount 3051C model code mean?

PositionWhat it fixesCommon codes
1Model3051C = Coplanar pressure transmitter
2Measurement typeD = differential, G = gage, A = absolute
3Pressure range0 to 5 (see the range table below)
4Transmitter outputA = 4–20 mA with HART, F = FOUNDATION Fieldbus, W = PROFIBUS PA, X = wireless, M = low-power 1–5 Vdc with HART
5Process flange type and material2 = Coplanar, stainless steel; 3 = Coplanar, cast C-276; 4 = Coplanar, Alloy 400; 5 = Coplanar, plated carbon steel; 0 = alternate process connection
6Isolating diaphragm2 = 316L stainless steel, 3 = Alloy C-276, 4 = Alloy 400, 5 = tantalum, 6 = gold-plated Alloy 400, 7 = gold-plated 316 stainless steel
7O-ringA = glass-filled PTFE, B = graphite-filled PTFE
8Sensor fill fluid1 = silicone, 2 = inert
9Housing material and conduit entryA = aluminium, ½–14 NPT; B = aluminium, M20; J = stainless steel, ½–14 NPT; K = stainless steel, M20
After 9OptionsBracket, display, hazardous-area approval, certificates and others

Which pressure range does each range code give?

Range code3051CD (differential)3051CG (gage)3051CA (absolute)
0−3 to 3 inH2O (7.5 mbar)not availablenot available
1−25 to 25 inH2O (62 mbar)−25 to 25 inH2O0 to 30 psia (2.07 bar)
2−250 to 250 inH2O (623 mbar)−250 to 250 inH2O0 to 150 psia (10.3 bar)
3−1000 to 1000 inH2O (2.49 bar)−393 to 1000 inH2O0 to 800 psia (55.2 bar)
4−300 to 300 psi (20.7 bar)−14.2 to 300 psi0 to 4000 psia (275.8 bar)
5−2000 to 2000 psi (137.9 bar)−14.2 to 2000 psinot available

The range code gives the limits of the sensor. The calibrated span of a delivered transmitter is set inside those limits and is shown separately on the nameplate.

Which option codes appear most often?

Option codeMeaning
B4Coplanar flange bracket, all stainless steel, for 2-inch pipe and panel mounting
M5LCD display
M4LCD display with local operator interface
E8ATEX flameproof and dust
I1ATEX intrinsic safety and dust
N1ATEX type n and dust
K8ATEX flameproof, intrinsic safety, type n and dust combined
E7IECEx flameproof and dust ignition-proof
I7IECEx intrinsic safety
E5FM explosion-proof and dust ignition-proof
I5FM intrinsically safe and nonincendive
Q4Calibration certificate
Q8Material traceability certification to EN 10204 3.1
QTSafety certified to IEC 61508
T1Transient protection terminal block

How is the example 3051CD2A22A1AB4 decoded?

The Emerson data sheet gives 3051CD 2 A 2 2 A 1 A B4 as a typical model number. Read position by position:

  • 3051C: Coplanar pressure transmitter.
  • D: differential pressure.
  • 2: range −250 to 250 inH2O.
  • A: 4–20 mA output with HART.
  • 2: Coplanar stainless steel process flange.
  • 2: 316L stainless steel isolating diaphragm.
  • A: glass-filled PTFE O-ring.
  • 1: silicone fill fluid.
  • A: aluminium housing with ½–14 NPT conduit entries.
  • B4: stainless steel mounting bracket.

Why does Alisios ask for the complete model code?

Two Rosemount 3051 transmitters that begin with the same characters can differ in range, output protocol, diaphragm material or hazardous-area approval. Each difference changes the part that is built and its price. Alisios quotes a Rosemount 3051 only from the complete model code on the nameplate, together with the serial number. The 3051T in-line and 3051L level transmitters follow the same principle with their own tables in the same data sheet.

Source: Emerson, Rosemount 3051 Pressure Transmitter product data sheet 00813-0100-4001. Emerson revises this data sheet and its option list. The current revision applies.

Have the equipment nameplate to hand?

Send a photo or the part number. We check the model and reply within 4 business hours.

Frequently asked questions

What does the D, G or A after 3051C mean?

It is the measurement type. 3051CD measures differential pressure, 3051CG measures gage pressure and 3051CA measures absolute pressure.

Is the range code the same as the calibrated range?

No. The range code fixes the limits of the sensor. The calibrated range is set within those limits when the transmitter is ordered or commissioned, and it is shown separately on the nameplate.

Which Rosemount 3051 option codes are ATEX and IECEx approvals?

In the Emerson data sheet, E8 is ATEX flameproof, I1 is ATEX intrinsic safety, N1 is ATEX type n, E7 is IECEx flameproof and I7 is IECEx intrinsic safety.

Get a quoteWhatsApp